Using ProLink
®
Software with Micro Motion
®
Transmitters
Configure Menu: Transmitter Information
continued
Pressure input for RFT9739
Config and Calib Registers
For custody transfer applications, security event registers enable the
user to determine whether the configuration or calibration of the
RFT9739 has been changed.
• Security event register display boxes will only be displayed if the
transmitter is a Version 3 RFT9739.
• Regardless of the transmitter security mode, the security event
registers count changes to the parameters listed in
Table 11-1
.
• Each security event register will count to 999, then roll over to
zero (0).
Table 11-1. Parameters that increment event registers
11.3 Pressure input for
RFT9739
The pressure input option allows the user to select the appropriate input
for either pressure compensation or viscosity measurement. Open the
Pressure Input drop-down list:
• Options in the drop-down list are described below.
• Based on the user's selection, a HART tag name or a 4-20 mA range
may need to be entered for the pressure device.
The RFT9739 accepts pressure input signals from a pressure
transmitter for viscosity measurement or pressure compensation. If the
signal from the pressure transmitter fails, both of the following occur:
• The RFT9739 continues to operate in a non-fault mode.
• A "Pressure Input Failure" message is shown on the transmitter
display (if it has one).
Primary viscosity
The RFT9739 polls for differential pressure values from a
HART-compatible DP cell as a primary master.
• The user must enter the HART tag name for the desired transmitter.
• The user must perform a viscosity calibration, as described in
Section 14.5
, page 148.
Secondary viscosity
The RFT9739 polls for differential pressure values from a
HART-compatible DP cell as a secondary master.
• The user must enter the HART tag name for the desired transmitter.
• The user must perform a viscosity calibration, as described in
Section 14.5
, page 148.
Configuration register see page: Calibration register see page:
Range values for milliamp outputs 113 Mass and volume measurement units 96
Variables assigned to milliamp outputs 112 Frequency/pulse output scaling factors 110
Variable assigned to control output 117 Flow calibration factor 73
Mass and volume flow cutoffs 102 Flowmeter zeroing 133
Internal damping 105 Density calibration factors 78, 80
Flow direction 104 Density calibration 138
Milliamp output trim 151 Pressure compensation factors 86
Master reset (see transmitter instruction manual) Meter factors for flow and density 131